Water characteristics associated with the occurrence of Legionella pneumophila in dental units

European Journal of Oral Sciences | 01/02/2000
Instalaciones Fontanería Gestión Higiene y desinfección hospitalaria Documentación Bibliografias Water characteristics associated with the occurrence of Legionella pneumophila in dental units


Eur J Oral Sci., 2000 Feb;108(1):22-8
Autores:

Zanetti, F.; Stampi, S.; De Luca, G.; Fateh-Moghadam, P.; Antonietta, M.; Sabattini, B.; Checchi, L.
Department of Medicine and Public Health, University of Bologna, Italy

Abstract:

This study evaluated the incidence of Legionella pneumophila in dental unit water samples and investigated how the occurrence of these bacteria may be related to some physical, chemical and bacteriological characteristics of the water. The samples were taken from the incoming tap water, oral rinsing cup, air-water syringe, ultrasonic scaler, and the turbine of 23 dental units of private and public institutions. Apart from L. pneumophila (serogroup 1 and 3) isolated in 22 out of the 101 (21.8%) water samples tested, two other species were found: L. bozemanii and L. dumoffii. The highest densities and frequency of L. pneumophila were observed in the water coming into the units and in the dental units of public institutions. A negative association between L. pneumophila and 36 degrees C and 22 degrees C heterotrophic total plate counts and other gram-negative bacteria was found. An inverse association between the concentration of L. pneumophila and water temperature was also observed. The values of pH and total hardness did not show any significant difference in the L. pneumophila-positive and -negative dental unit waters. Finally, the chemical oxygen demand (COD) and residual chlorine were found to correlate positively with L. pneumophila.
Zanetti, F. ... [et al.]

Para poder escribir un comentario debe iniciar sesión o darse de alta en el portal.

Patrocinadores Platino

Patrocinadores Plata